97

Advanced Operations

6.3.3.5 Function set to WARNING working mode

WARNING

WARNING LIST

Default value: Disabled
Sets the monitoring of the internal warning bits of the selected source.
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.

ERROR LIST

Default value: Disabled
Sets the monitoring of the internal error bits of the selected source.
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.

PROCESS EVENT

Default value: Disabled
Sets the monitoring of the internal process event bits of the selected source.
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.

SET TRANSFER

Default value: DE-ENERGIZED
Sets the status of the relay (ENERGIZED/DE-ENERGIZED) if some or all conditions (this means
warning, error or process event bits) are detected in the selected source or if the source is
missing.

ON DELAY
(0 s

999 s)

Default value: 5 seconds
Sets a delay time for the relay to turn on.

OFF DELAY
(0 s

999 s)

Default value: 5 seconds
Sets a delay time for the relay to turn off.