4 modbus card connections, Installation 3.6.4 modbus card connections – Hach sc1000 User Manual

Page 36

Advertising
background image

34

Installation

3.6.4 Modbus card connections

Modbus RS485 (YAB021) and Modbus RS232 (YAB047) are available. For more detailed
information refer to the bus system manual.

To make a Modbus card connection:

1. Remove power from the instrument. Remove the probe module cover.

2. Connect the Modbus card to the appropriate slot (

Figure 18

). Use a magnetic

screwdriver to secure the four screws to the card.

3. Install the card connector to the appropriate connection on the main circuit board

(

Figure 17

).

4. Feed the cable through the base of the module and properly prepare and insert each

wire into the terminal according to

Figure 24

/

Table 7

and

Figure 25

/

Table 8

.

5. Write the serial number from the rating plate on the supplied sticker and attach it to the

main high voltage barrier (

Figure 18

).

6. Install the probe module cover.

After installation and connection of a plug-in expansion card, the card must be configured
to the system. For Modbus card setup instructions, refer to

section 6.3.4.2 on page 111

.

)

Figure 24 Modbus RS485 (YAB021) card connections

1

Card (Reverse side)

3

Jumper 1&2 unplugged for full duplex (4-wire)

2

Jumper 1&2 plugged for half duplex (2-wire)

4

Terminal block (Refer to

Table 7

for terminal

assignments)

Table 7 Modbus RS485 card (YAB021) terminal assignments

Terminal

Modbus RS485 designation with 4 wires

Modbus RS485 designation with 2 wires

1

Not Used

Not Used

2

Not Used

Not Used

3

Output –

4

Output +

+

5

Input –

6

Input +

+

7

Shield (connected to Protective Earth)

Shield (connected to Protective Earth)