2 error and status register, Standard operations – Hach sc1000 User Manual

Page 66

Advertising
background image

64

Standard Operations

7. Select a tag and press the

ENTER

button. The new tag is added to the Telegram list.

Select a tag and press the

UP

and

DOWN

button to move the position of the tag

(

Figure 47

and

Table 14

).

8. Repeat the steps to add further devices and tags.

9. Press the

ENTER

button to store the Profibus configuration.

5.12.2 Error and status register

Note: ERROR and STATUS definitions are valid for all sc probes.

Figure 47 Profibus/Modbus configuration menu

Telegram list with new tag

Table 14 Telegram list—Column description

Column

Description

1

Profibus: Data position in the configured Profibus slave (in 2 byte words)
Modbus: Data position in the configured Modbus slave
This slave contains holding registers beginning at 40001.
Example: “0” means register 40001 or “11” means register 40012.

2

Tag name to identify the configured data.

3

Data type
float=floating point value
int=integer numbers
sel=integer value resulting from a enum (selection) list

4

Data status
r=data are read only
r/w=read/write (for DPV0 only)

Table 15 Error register

Bit

Error

Description

0

Measurement calibration error

An error has occurred during the last calibration.

1

Electronic adjustment error

An error has occurred during the last electronic
calibration.

2

Cleaning error

The last cleaning cycle failed.

3

Measuring module error

A failure in the measuring module has been detected.

4

System re-initialization error

Some settings have been detected inconsistent and set
to factory defaults.

Advertising