Installation – Hach sc1000 User Manual

Page 31

Advertising
background image

29

Installation

To make a relay card connection
For instruments not equipped with a relay card, do the steps listed below to make relay
card connections.

1. Remove power from the instrument. Remove the probe module cover.

2. Remove the screws on the plastic relay cover. Remove the plastic cover.

3. Connect the relay card to the appropriate slot (

Figure 18

). Use a magnetic

screwdriver to secure the four phillips-head screws to the card. It is easier to connect
the cards with the module in its normal vertical mounted position, rather than laid
horizontally on a bench.

4. Install the card connector to the appropriate connection on the main circuit board

(

Figure 17

).

5. Feed the cable through the base of the module and properly prepare and insert each

wire (

Figure 19

) into the terminal according to

Figure 20

/

Table 3

and

Figure 21

/

Table

4

. Tug gently after each insertion to make sure that the connection is secure.

6. Write the serial number from the rating plate on the supplied sticker and attach it to

the main high voltage barrier (

Figure 18

). This serial number is the same internal

address of the card on the network.

7. Install the relay and probe module cover.

After installation and connection of a plug-in expansion card, the card must be configured
to the system. For relay card setup instructions, refer to

section 6.3.3 on page 87

.

Figure 19 Proper wire preparation and insertion

1

Strip ¼-in (64 mm) of insulation.

2

Seat insulation against connector with no bare wire
exposed.

Figure 20 Relay card (YAB022, normally closed)

1

Terminal Block–Refer to

Table 3

for terminal assignments.

Advertising