Advanced operations – Hach sc1000 User Manual

Page 88

Advertising
background image

86

Advanced Operations

Figure 53

shows the output value depending on the input current, the set low value and

the set high value with an input range of 4

20 mA.

The output value (OV) is determined according to formula

(4)

:

(4)

where:
OV=Output value
IC=Input current
LV=Low value
HV=High value

Figure 53 Output value with an input range of 4

20 mA

1

Output value (concentration) (y-axis)

5

4 mA

2

OV=f(IC)

6

0 mA

3

Input current (x-axis)

7

Low value (LV)

4

20 mA

8

High value (HV)

OV

f(IC)

HV-LV

16mA

-----------------

(IC-4mA)

×

LV

+

=

=

Advertising