A.5 attach the external output module – Hach sc1000 User Manual

Page 147

Advertising
background image

145

A.5 Attach the external output module

The output module provides two outputs, 0–20 mA/4–20 mA, 500 Ohm.

1. Remove power and disconnect all cable connections to the module.

2. Place the External Output Module on the DIN rail to the right of the base module and

slide firmly against the base module (or other connected modules).

3. Make the appropriate connections as shown in

Figure 78

and

Table 51

.

4. Connect cables to the module and reconnect power from the instrument.

Figure 78 External output module

Table 51 External Output Module Terminal Assignments

Terminal

Assignment

Description

1–4

Not used

5

Analog output 1 max. 500 Ohm

6

Shield

7

+

8

Not used

9

Analog output 2 max 500 Ohm

10

+

11

Shield

12–16

Not used

Advertising