79

Advanced Operations

SYSTEM SETUP
OUTPUT SETUP
mA OUTPUT INT/EXT

Select OUTPUT card 1,2,3 or 4

SELECT SOURCE

Default value: No source
Selects a probe or creates a formula which delivers the process value that is processed by the
current output card.

SET PARAMETER

Default value: No Parameter
Selects a parameter of the selected source.

DATA VIEW

Default value: INPUT VALUE
Sets the displayed and logged measured value.

INPUT VALUE

Displays the process value which is read from the selected source after it has been processed
by the internal formula parser (if prompted)

CURRENT

Displays the calculated output current

SET FUNCTION

Default value: LINEAR CONTROL

LINEAR CONTROL

Tracks the measurement value.

PID CONTROL

Sets the sc1000 controller as a PID controller.

SET TRANSFER

Default value: 10 mA
Sets the substitute value for the output current in case the selected source reports an internal
error, is disconnected from the system or its output mode is set to 'Transfer value'.

ON ERROR MODE

Default value: SET TRANSFER
Sets the sc1000 controller reaction if an internal error occurs.

HOLD

The current output card operates constantly with the last valid value read from the selected
source.

SET TRANSFER

The current output card uses a substitute value for the output current.

SET MODE

Default value: DIRECT
Defines the point in time when the PID controller increases the Output current

DIRECT

The SNAP SHOT value is lower than the SETPOINT and vice versa.

REVERSE

The SNAP SHOT value is higher than the SETPOINT and vice versa.

SET FILTER

Sets the record time (in seconds)
The output current is based on the average of recorded values during a period of time. The
period set in this menu.

SCALE 0 mA/4 mA

Default value: 0-20 mA
Sets the output current range to 0–20 mA or 4–20 mA.

SET HIGH VALUE

Default value: 20
Sets the selected source value when the output current shall be 20 mA.

SET LOW VALUE

Default value: 0
Sets the selected source value when the output current shall be 0 mA (Scale is 0-20 mA)
respectively 4 mA (Scale is 4–20 mA).

MAXIMUM

Default value: 20 mA
Sets an upper limit for the possible output current value.
This menu item is displayed if SET FUNCTION is set to PID CONTROL.

MINIMUM

Default value: 0 mA
Sets the lower limit for the output current.
This menu item is displayed if SET FUNCTION is set to PID CONTROL.

SET SETPOINT

Default value: 10
Sets the process value
The PID controller tries to adjust to this process value.