Advertising
background image

66

Standard Operations

5.12.3 Profibus/Modbus configuration example

Table 17

and

Table 18

show a Profibus/Modbus configuration example.

For more information concerning Profibus configuration settings, refer to

section 6.3.4.1

on page 109

.

For more information concerning Modbus configuration settings, refer to

section 6.3.4.2

on page 111

.

Table 17 Profibus configuration example

Profibus
address

Slave

Byte

Device

Data name

5

Configured Slave

1,2

AMTAX SC

ERROR

3,4

STATUS

5,6,7,8

CUVETTE TEMP

9,10,11,12

MEASURED VALUE 1

13,14

mA INPUT INT

ERROR

15,16,

STATUS

17,18,19,20

INPUT CURRENT 1

21,22

DIGITAL INPUT 2

23,24,25,26

OUTPUT VALUE 3

27,28

DIGITAL INPUIT 4

Table 18 Modbus configuration example with virtual slaves

Modbus
address

Slave

Register

Device

Data name

5

Configured Slave

40001

AMTAX SC

ERROR

40002

STATUS

40003

CUVETTE TEMP

40005

MEASURED VALUE 1

40007

mA INPUT INT

ERROR

40008

STATUS

40009

INPUT CURRENT 1

40011

DIGITAL INPUT 2

40012

OUTPUT VALUE 3

40014

DIGITAL INPUIT 4

6

First virtual slave
(AMTAX SC)

40001

AMTAX SC (complete)

See AMTAX SC profile

40002

See AMTAX SC profile

...

See AMTAX SC profile

7

Second virtual slave
(mA INPUT INT)

40001

mA INPUT INT (complete)

See mA INPUT INT profile

40002

See mA INPUT INT profile

...

See mA INPUT INT profile