65

Standard Operations

5

Hardware error

Hardware error detected.

6

Internal communication error

A communication failure within the device has been
detected.

7

Humidity error

Excessive humidity has been detected.

8

Temperature error

Temperature within the device exceeds specified limit.

10

Sample warning

Some action is required with the sample system.

11

Questionable calibration warning

The last calibration was of questionable accuracy.

12

Questionable measurement warning

One or more of the measurements of the device are of
questionable accuracy (Quality bad or out of range).

13

Safety warning

A condition has been detected which may result in a
safety hazard.

14

Reagent warning

Some action is required with the reagent system.

15

Maintenance required warning

Maintenance is required on this device.

Table 15 Error register

Bit

Error

Description

Table 16 Status register

Status 1

Bit

Status 1

Description

0

Calibration in progress

The device is in calibration mode. The measurements
may not be valid.

1

Cleaning in progress

The device is in cleaning mode. The measurements may
not be valid.

2

Service/Maintenance menu

The device is in service or maintenance mode. The
measurements may not be valid.

3

Common error

Device recognized an error, see

Table 15

for more

detailed information

4

Measurement 0 Quality bad

Precision of measurement is out of specified limits.

5

Measurement low limit

Measurement is below specified range.

6

Measurement high limit

Measurement is above specified range.

7

Measurement 1Quality bad

Measurement is below specified range.

8

Measurement 1 low limit

Measurement is above specified range.

9

Measurement 1 high limit

Measurement is below specified range.

10

Measurement 2 Quality bad

Measurement is above specified range.

11

Measurement 2 low limit

Measurement is below specified range.

12

Measurement 2 high limit

Measurement is above specified range.

13

Measurement 3 Quality bad

Measurement is below specified range.

14

Measurement 3 low limit

Measurement is above specified range.

15

Measurement 3 high limit

Measurement is below specified range.