2 function set to alarm working mode, Advanced operations – Hach sc1000 User Manual

Page 90

Advertising
background image

88

Advanced Operations

6.3.3.2 Function set to ALARM working mode

FEEDER CONTROL

Operates in response to the measured parameter. Can be set for phasing, set point, deadband,
overfeed timer, and ON/OFF delay.

2 POINT CONTROL

Operates in response to the measured parameter using two setpoints.

WARNING

Activated when the analyzer detects a probe warning. Indicates warning and error condition of
selected probes.

PWM Control

Allows the relay to provide a pulse width modulated output.

FREQ. CONTROL

Allows the relay to cycle at a frequency between the minimum pulses per minute and maximum
pulses per minute.

TIMER

Allows the relay to switch at certain times independent of any process value

SYSTEM ERROR

Indicates if a probe in the system has an internal error or warning

INPUT VALUE

The process value read from the selected source after it has been processed by the internal
formula parser (if prompted).

LOG INTERVAL

Default value: OFF
Sets the interval for logging the displayed value into the data logger.
Options: OFF, 5min, 10min, 15min, 20min, 30min

SYSTEM SETUP
RELAY
RELAY INT/EXT

ALARM

SET TRANSFER

Default value: DE-ENERGIZED
Sets the status of the relay (ENERGIZED/DE-ENERGIZED) if error conditions are detected in
the selected source or if the source is missing.

PHASE

Default value: DIRECT
Decides if the relay is turned on or off when the process value is leaving the controlled band.

DIRECT

Relay is turned on when leaving the controlled band

REVERSE

Relay is turned off when leaving the controlled band

HIGH ALARM

Default value: 15
Sets the maximum of the controlled band in the selected parameter unit.

LOW ALARM

Default value: 5
Sets the minimum of the controlled band in the selected parameter unit.

HIGH DEADBAND

Default value: 1
Sets the hysteresis value used at the upper limit.

LOW DEADBAND

Default value: 1
Sets the hysteresis value used at the lower limit

ON DELAY
(0 s–999 s)

Default value: 5 seconds
Sets a delay time for the relay to turn on.

OFF DELAY
(0 s–999 s)

Default value: 5 seconds
Sets a delay time for the relay to turn off.