Chapter 5 - Repairs & Replacements

Fuse Holder,

(Floor Temperature,

Left & Right)

P/N 853671

There are two fuse/fuse holder assemblies on the electronics
compartment door, one each for left floor temperature and right
floor temperature. The fuse holders are identical and are removed
and installed the same way.

CAUTION: Before proceeding, make sure the unit is
Off, and the electric power cord un-plugged.

Tool and Supplies Required

„

11/16-in. wrench

„

Markers, masking tape, tags, or other means to mark wires

Removal

1.

Make sure that the main on/off switch is
Off and the amber light in the switch is out.

2.

Unplug the power cord from the electrical
outlet.

3.

Open the electronics compartment (refer
to Page 57).

4.

Mark both wires on the rear of the fuse
holder so that you can re-install them
correctly.

5.

Dis-connect both wires from the fuse holder. Note: There are
two terminals on the fuse holder: one on the side and one on
the rear (Figure 121).

6.

With a 11/16-in. wrench, unscrew and remove the nut holding
the fuse holder to the door.

7.

Pull the fuse holder out through the front of the door.

8.

If the fuse is still good, you can remove the it from the fuse
holder and re-use it.

Figure 121. Inside View of Fuse Holder