Chapter 5 - Repairs & Replacements

Audible Alarm

P/N 853458

11.

Mount the thermostat and bracket to the
partition wall (Figure 124).

12.

Neatly coil up the excess capillary tubing
and bind it to the adhesive tie mount with a
small wire tie (Figure 129).

13.

Replace the Sodalime pan.

14.

Close and secure the filter/fans compart-
ment panel (Page 58).

15.

Close and secure the electronics compart-
ment door (Page 57).

16.

Plug the power cord into its wall outlet.

CAUTION: Before proceeding, make sure the unit is
Off, and the electric power cord un-plugged.

Tool and Supplies Required

„

Phillips screwdriver

„

Markers, masking tape, tags, or other means to mark wires

Removal

1.

Make sure that the main on/off switch is Off and the
amber light in the switch is out.

2.

Unplug the power cord from the electrical outlet.

3.

Open the electronics compartment (refer to Page 57).

4.

Using tags, masking tape, or any other convenient means,
tag the wires with the identification of the terminal to which
they attach (Figure 123). Failure to mark the wires will
make it difficult to connect them correctly later.

5.

With a Phillips screwdriver, remove all wires from the rear
of the audible alarm.

6.

Close the electronics compartment door.

Figure 132. Thermostat Knock-out

Knock-out

Wire

Terminals