Excessive carbon dioxide in the animal compartment.

First: The Sodalime supply in the filter/fans compartment is
probably exhausted. Replace the Sodalime with fresh material.
Refer to Refilling the CO

2

Filter on Page 65.

Second: Make sure the circulating fans in the filter/fans
compartment are working. Refer to Circulating fan does not work
on Page 149.

Third: Make sure that the air inlet grill and the air outlet grill
(Figure 3 on Page 9) are clear and unobstructed.

Fourth: Make sure that your set-point setting on the oxygen
concentration meter is correct. Refer to Using Oxygen on
Page 51.

Fifth: Make sure that the flow rate on the oxygen flow meter is
correct. Refer to Controlling Oxygen Flow Rate on Page 56.

Sixth: Check your oxygen supply and pressure regulator setting.

Seventh: Refer to Oxygen flow is inadequate on Page 141.

Remedial Action