57

Owner’s Manual - Regal Intensive Care Unit

Accessing the

Electronics

Compartment

Figure 60. Electronics Compartment
Access Knob

Panel

Access

Knob

Electronics

Compartment

(shown partly

open)

Figure 61. Electronics Compartment
Panel Interior

To close

panel, latch

arm must

point down

Inside of

Access

Knob

To close and lock the compartment, check the inside of the panel
and verify that the latch arm is pointed down (Figure 61). Then,
swing the panel closed and turn the panel access knob one-quarter
turn clockwise.

CAUTION: Opening the electronics compartment
door exposes electrical components inside. Use
caution around electric wiring and components to
prevent injury to yourself and/or damage to the
equipment. Before working in the electronics
compartment, press the main on/off switch to Off, and
un-plug the electric power cord unless electric power
is required for servicing.

To gain access to the electronics compartment, turn the panel
access knob one-quarter turn counter-clockwise and pull the panel
open (Figure 60). The panel is hinged on the right side and opens
like a door.