Oxygen flow is inadequate – Suburban Mfg REGAL INTENSIVE CARE UNIT 12019-01-NGFREP User Manual

Page 153

Advertising
background image

Form No. 702718 Rev. A March, 2005

141

Owner’s Manual - Regal Intensive Care Unit

Oxygen flow is inadequate.

First: Check your oxygen supply. Is the tank empty or is the
pressure regulator improperly adjusted?

Second: Make sure that all four panel access knobs on the filter/
fans compartment are firmly locked. Failure to lock all four of these
knobs will result in oxygen leakage through the air grills between the
animal and filter/fans compartments, and consequent loss of oxygen
past the loose panel.

Third: Is the oxygen flow meter adjusted properly? Refer to
Controlling Oxygen Flow Rate on Page 56.

Fourth: Is the set point on the oxygen concentration meter
correctly set? Refer to Using Oxygen on Page 51.

CAUTION: If oxygen pressure, flow rate, and
concentration are correctly set, and the oxygen flow
is still low, you probably have a leak in the system.
Turn the ICU off and observe all oxygen safety
measures. Refer to Oxygen Warnings
on Page 3.

Fifth: Have your oxygen supply service check your supply lines
and connections for leaks.

CAUTION: Before working on the electrical system,
make sure the unit is Off and the electric power cord
un-plugged.

Sixth: Open the electronics compartment and check all oxygen
lines. Check for loose connections, and damaged or kinked oxygen
lines.

Seventh: Your solenoid valve may be defective. Order P/N
853440 and refer to Page 118 for replacement instructions.

Eighth: Your oxygen flow meter may be defective. Order
P/N 853456 and refer to Page 120 for replacement instructions.

Remedial Action