Panel, filter/fans, complete – Suburban Mfg REGAL INTENSIVE CARE UNIT 12019-01-NGFREP User Manual

Page 140

Advertising
background image

128

SSCI - Suburban Surgical Company, Inc.

Chapter 5 - Repairs & Replacements

Panel, Filter/Fans,

Complete

P/N 206874

Procedure

1.

Turn all four panel access knobs (Figure 149) one-quarter
turn counter-clockwise and pull the old panel completely
off the Intensive Care Unit.

2.

Look at the rear of the new access panel (Figure 150) and
make sure that the locking arms on all four access knobs
are vertical. Note: The locking arms on the right should
point down, and that the arms on the left should point up.

3.

Seat the panel in place over the filter/fans compartment
opening, and turn all four panel knobs one-quarter turn
clockwise to lock the panel in place. Note: It is a good
idea to tug gently on one of the access knobs to be sure the
panel is secure.

Figure 149. Filter/Fans Compartment
Access Knobs

Access

Knobs

Note: If the digital readout from the old filter/fans compartment
panel is working OK, it can be dis-mounted from the panel and
used in some other way, or kept as a spare.

Figure 150. Inside of Filter/Fans
Compartment Panel

To close

panel, latch

arms on this

side must

point down.

To close

panel, latch

arms on this

side must

point up.