SSCI - Suburban Surgical Company, Inc.

Chapter 5 - Repairs & Replacements

Parts Replacement
Procedures

Compartment Door

Left - P/N 205994

Right - P/N 205995

The two animal compartment doors on the Intensive Care Unit are
similar and are removed and installed in the same way.

Procedure

1.

Open the door.

2.

Lift the complete door up until the upper halves of both
door hinges come free of the lower halves (Figure 82).

3.

Holding the door firmly, align the pins in the upper halves
of both door hinges with the holes in the lower halves, and
gently lower the door onto the hinges (Figure 82).

4.

Gently swing the door open and closed several times to
make sure it moves freely.

5.

Close and latch the door.

Figure 82. Compartment Door Hinges