Chapter 3 - Installation & Setup

Figure 39. Oxygen Concentration Meter
and Calibration Knob

Black Calibration

Needle

Calibration

Knob

Figure 38. Fuel Cell and Oxygen Sensor
Assembly

Sensor

End
Cap

Fuel

Cell

Oxygen

Sensor

Body

This end of Fuel Cell goes into sensor body

„

If the oxygen sensor is mounted in the large
black plastic ring, you have a New Style
Intensive Care Unit (produced after
January 1, 2005). Proceed to New Style
Units
on Page 37.

„

If this large black ring is not present, you
have an Original Style Intensive Care Unit
(produced before January 1, 2005). Follow
the instructions under Original Style
Units
below.

Original Style Units:

1.

In the electronics compartment (Figure 36)
twist the sensor slightly, and pull it out of
the compartment wall. Note: Do not try
to detach the spiral cord from the sensor.

2.

Unscrew the end cap from the oxygen sensor (Figure 38).

3.

On the fuel cell, remove the shorting button - refer to the
instructions packed with the fuel cell.

Note: Watch the oxygen concentration meter (Figure 39) as you
perform Step 4. The black calibration needle moves to the right,
then settles down.

4.

Place the fuel cell into the oxygen sensor body. The end
of the fuel cell with the metal disk goes in first.