60

SSCI - Suburban Surgical Company, Inc.

Chapter 4 - Operation and Care

Re-setting the

Circuit Breakers

Two 10 amp. circuit breakers are provided on the face of the
electronics compartment (Figure 68). Circuit breakers are safety

devices that shut down electrical power to the unit
should an overload or short circuit occur.

The left breaker (viewed from the front) protects
the cooling system. The right breaker protects the
remainder of the unit.

When operating normally and not “tripped”, the white
button in the center of the circuit breaker is level with
the black outer ring of the breaker. When “tripped” the
white button projects about 1/8-in. out from the black
outer ring.

Should a circuit breaker trip, wait at least 10-seconds,
then press the white button in to reset it.

Occasional tripping of a circuit breaker is not a cause for concern,
however, if circuit breakers trip frequently, it may be a sign of
problems in the unit’s electrical system. The unit should be checked
and repairs made as needed.

3.

On the rear of the digital readout, click the
switch labeled F°- C° (Figure 67) to the
desired setting.

4.

Replace and secure the compartment
panel.

Figure 68. Circuit Breakers

Figure 67. Rear of the Digital Readout

F° - C° Switch