SSCI - Suburban Surgical Company, Inc.

Chapter 3 - Installation & Setup

Checking the

Oxygen System

Note: You can perform this test even though oxygen has not yet
been hooked up.

1.

Push the main on/off switch (Figure 40) to On. The amber
light in the switch comes on.

2.

Push the oxygen on/off switch to On (Figure 44). The
amber light in the switch comes on. If the black calibration
needle in the oxygen concentration meter (Figure 45) is to the
left of the red set-point needle, the red system light will also
come on.

3.

If the black needle is to the right of the red needle, use the
set-point adjustment knob to place the red needle to the
right of the black needle. The red system light comes on.

4.

Push the main on/off switch to Off. The light in the switch
goes out.

5.

Return the black calibration needle to the 20.5% setting.

Figure 45. Oxygen Concentration Meter
and Controls

Red Set-point

Needle

Black Calibration

Needle

Calibration

Knob

Set-point

Adjustment Knob

Figure 44. Oxygen On/Off Switch and
System Light

On/Off
Switch

System

Light