Owner’s Manual - Regal Intensive Care Unit

The alarm system on the Regal Intensive Care Unit warns you if
there is a problem with the oxygen supply or concentration.

1.

Open the electronics compartment (refer to Page 57).

2.

Set the alarm time delay relay to 30-seconds (Figure 46).

3.

With the calibration knob (Figure 45), adjust the black
calibration needle in the oxygen concentration meter to
20.5% (the percentage of oxygen in atmospheric air).

4.

With the set-point adjustment knob, position the red set-
point needle to about 30%.

5.

Push the main on/off switch (Figure 40) to On. The amber
light in the switch comes on.

6.

Push the oxygen on/off switch (Figure 44) to On. The amber
light in the switch comes on.

7.

Push the alarm on/off switch (Figure 47) to On. The amber
light in the switch comes on.

Checking the

Alarm Function

Figure 47. Alarm Controls

On/Off
Switch

System

Light

8.

After 30-seconds, since oxygen is not hooked-up, the
system senses that the oxygen concentration is inadequate.
The audible alarm sounds and the red system light goes on.

9.

Move the red set-point needle to 10%. As the red needle
passes the black needle, the audible alarm stops. Both the
alarm and oxygen system lights go out.

Figure 46. Alarm Time Delay Relay

Delay Adjustment Knob