Isolation and Digital Isolators

USB-6215/6218 devices are isolated data acquisition devices. As shown in
Figure 11-1, the analog input, analog output, counters, PFI/static DI, and
PFI/static DO circuitry, and digital routing and clock generation are
referenced to an isolated ground. The bus interface circuitry is referenced
to a non-isolated ground. Refer to Table 11-1 for an example of the
symbols for isolated ground and non-isolated ground.

Figure 11-1. General NI 621x Block Diagram

The non-isolated ground is connected to the chassis ground of the PC.

Each isolated ground is not connected to the chassis ground of the PC. The
isolated ground can be at a higher or lower voltage relative to the
non-isolated ground. All analog measurements are made relative to the
isolated ground signal.