Figure 9-9. Buffered Period Measurement

Note that if you are using an external signal as the Source, at least one
Source pulse should occur between each active edge of the Gate signal.
This condition ensures that correct values are returned by the counter. If this
condition is not met, the counter returns a zero. Refer to the

Duplicate

Count Prevention

section for more information.

For information about connecting counter signals, refer to the

Default

Counter/Timer Pinouts

section.

Table 9-1. Time N Descriptions

t

0

At t

0

, the counter is armed. No measurements are taken until

the counter is armed.

t

1

The rising edge of Gate indicates the beginning of the first
period to measure. The counter begins counting rising edges of
Source.

t

2

The rising edge of Gate indicates the end of the first period. The
USB M Series device stores the counter value in the buffer.

t

3

The rising edge of Gate indicates the end of the second period.
The USB M Series device stores the counter value in the buffer.

SOURCE

GATE

Counter Value

Buffer

1

3

3

3

1

1

2

2

3
3

3

Counter Armed

Time N

3

t

0

t

1

t

2

t

3