Floating signal sources -8, Floating signal sources – National Instruments NI USB-621x User Manual

Page 58

Advertising
background image

Chapter 5

Connecting AI Signals on the USB-6210/6211 Devices

NI USB-621x User Manual

5-8

ni.com

inputs; be aware that there is some gain error from loading down the source,
as shown in Figure 5-4.

Figure 5-4. Differential Connections for AC Coupled Floating Sources

with Balanced Bias Resistors

Using Non-Referenced Single-Ended (NRSE) Connections for Floating
Signal Sources

It is important to connect the negative lead of a floating signals source to
AI GND (either directly or through a resistor). Otherwise the source may
float out of the valid input range of the NI-PGIA and the DAQ device
returns erroneous data.

Figure 5-5 shows a floating source connected to the DAQ device in NRSE
mode.

Figure 5-5. NRSE Connections for Floating Signal Sources

All of the bias resistor configurations discussed in the

Using Differential

Connections for Floating Signal Sources

section apply to the NRSE bias

resistors as well. Replace AI– with AI SENSE in Figures 5-1, 5-2, 5-3,
and 5-4 for configurations with zero to two bias resistors. The noise
rejection of NRSE mode is better than RSE mode because the AI SENSE
connection is made remotely near the source. However, the noise rejection
of NRSE mode is worse than DIFF mode because the AI SENSE

+

AI GND

V

s

AC Coupled

Floating

Signal

Source

AI+

AI–

AI SENSE

AC Coupling

+

AI GND

R

AI SENSE

AI

V

s

Floating

Signal

Source

Advertising