Chapter 5

Connecting AI Signals on the USB-6210/6211 Devices

NI USB-621x User Manual

5-12

ni.com

Using Differential Connections for Ground-Referenced Signal Sources

Figure 5-7 shows how to connect a ground-referenced signal source to the
USB-6210/6211 device configured in DIFF mode.

Figure 5-7. Differential Connections for Ground-Referenced Signal Sources

With this type of connection, the NI-PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the signal
source and the device ground, shown as V

cm

in the figure.

AI+ and AI– must both remain within ±11 V of AI GND.

M Series Device Configured in DIFF Mode

PGIA

+

+

+

+

V

cm

V

s

Ground-

Referenced

Signal

Source

Common-

Mode

Noise and

Ground

Potential

AI GND

AI SENSE

Input Multiplexers

V

m

Measured

Voltage

Instrumentation

Amplifier

AI+

AI–

I/O Connector