Figure 4-10. ai/sampleclock too fast, Figure 4-11. ai/convertclock too fast, Gures 4-10, 4-11 – National Instruments NI USB-621x User Manual
Page 45
Advertising

Chapter 4
Analog Input
© National Instruments Corporation
4-19
Figure 4-10. ai/SampleClock Too Fast
Figure 4-11. ai/ConvertClock Too Fast
Figure 4-12. ai/SampleClock and ai/ConvertClock Improperly Matched
• Sample Clock Too Fast for Convert Clock
ai/SampleClock
ai/ConvertClock
Sample #1
Sample #2
Sample #3
Channel Measured
1 2 3
0
1 2 3
0 1
2
3
0
• Sample Clock Pulses are Gated Off
• Convert Clock Too Fast for Sample Clock
• Convert Clock Pulses are Gated Off
ai/SampleClock
ai/ConvertClock
Sample #1
Sample #2
Sample #3
1 2 3
0
1 2 3
0
1 2 3
0
Channel Measured
• Improperly Matched Sample Clock and Convert Clock
• Leads to Aperiodic Sampling
ai/SampleClock
ai/ConvertClock
Sample #1
Sample #2
Sample #3
1
2
3
0
0
Channel Measured
1
2
3
0
Advertising