Chapter 4

Analog Input

NI USB-621x User Manual

4-12

ni.com

Figure 4-4. Analog Input Timing Options

M Series devices use ai/SampleClock and ai/ConvertClock to perform
interval sampling. As Figure 4-5 shows, ai/SampleClock controls the
sample period, which is determined by the following equation:

1/Sample Period = Sample Rate

Figure 4-5. Interval Sampling

ai/ConvertClock controls the Convert Period, which is determined by the
following equation:

1/Convert Period = Convert Rate

20 MHz Timebase

100 kHz Timebase

Programmable

Clock

Divider

Programmable

Clock

Divider

ai/SampleClock

Timebase

ai/ConvertClock

Timebase

Analog Comparison Event

Ctr

n

Internal Output

SW Pulse

Ctr

n

Internal Output

ai/ConvertClock

ai/SampleClock

Channel 0

Channel 1

Convert Period

Sample Period