Non-Referenced Single-Ended (NRSE) Measurements

To reach a compromise between RSE and differential measurements, you
can use an NRSE measurement configuration. This configuration allows
for a remote sense for the negative (–) input of the programmable gain
instrumentation amplifier (PGIA) that is shared among all channels
configured for NRSE mode. The behavior of this configuration is similar to
the behavior of RSE connections but it provides improved noise rejection.
Refer to Figure 6-3 for an illustration of an NRSE connection
configuration.

Figure 6-3. Connecting a Device to a USB-6215/6218 Using NRSE Connections

In the NRSE connection configuration, each input channel is measured
with respect to AI SENSE.

AI1

AI0

AI GND

MUX

ADC

PGIA

AISENSE