Analog output timing signals, Figure 7-3. analog output timing options, Ao start trigger signal – National Instruments NI USB-621x User Manual

Page 72: Using a digital source, Analog output timing signals -5, Ao start trigger signal -5, Using a digital source -5

Advertising
background image

Chapter 7

Analog Output

© National Instruments Corporation

7-5

NI USB-621x User Manual

Analog Output Timing Signals

Figure 7-3 summarizes all of the timing options provided by the analog
output timing engine.

Figure 7-3. Analog Output Timing Options

USB M Series devices feature the following AO (waveform generation)
timing signals.

•

AO Start Trigger Signal

•

AO Pause Trigger Signal

•

AO Sample Clock Signal

•

AO Sample Clock Timebase Signal

AO Start Trigger Signal

Use the AO Start Trigger (ao/StartTrigger) signal to initiate a waveform
generation. If you do not use triggers, you can begin a generation with a
software command.

Using a Digital Source

To use ao/StartTrigger, specify a source and an edge. The source can be one
of the following signals:

•

A software pulse

•

PFI <0..3>, PFI <8..11>

•

ai/StartTrigger

20 MHz Timebase

100 kHz Timebase

Programmable

Clock

Divider

ao/SampleClock

Timebase

Ctr

n Internal Output

ao/SampleClock

PFI

PFI

SampleClock

Timebase Divisor

Advertising