Figure 4-2. ni-pgia, Table 4-3. signals routed to the ni-pgia – National Instruments NI USB-621x User Manual

Page 30

Advertising
background image

Chapter 4

Analog Input

NI USB-621x User Manual

4-4

ni.com

The AI ground-reference setting determines how you should connect your
AI signals to the NI 621x device. Refer to Chapter 5,

Connecting AI Signals

on the USB-6210/6211 Devices

, section for more information.

Ground-reference settings are programmed on a per-channel basis. For
example, you might configure the device to scan 12 channels—four
differentially-configured channels and eight single-ended channels.

NI 621x devices implement the different analog input ground-reference
settings by routing different signals to the NI-PGIA. The NI-PGIA is a
differential amplifier. That is, the NI-PGIA amplifies (or attenuates) the
difference in voltage between its two inputs. The NI-PGIA drives the ADC
with this amplified voltage. The amount of amplification (the gain), is
determined by the analog input range, as shown in Figure 4-2.

Figure 4-2. NI-PGIA

Table 4-3 shows how signals are routed to the NI-PGIA.

For differential measurements, AI 0 and AI 8 are the positive and negative
inputs of differential analog input channel 0. For a complete list of signal

Table 4-3. Signals Routed to the NI-PGIA

AI Ground-Reference

Settings

Signals Routed to the Positive

Input of the NI-PGIA (V

in+

)

Signals Routed to the Negative

Input of the NI-PGIA (V

in–

)

RSE

AI <0..31>

AI GND

NRSE

AI <0..31>

AI SENSE

DIFF

AI <0..7>

AI <8..15>

AI <16..23>

AI <24..31>

V

in+

V

m

= [V

in+

– V

in–

]

× Gain

V

m

V

in–

PGIA

+

Measured

Voltage

Instrumentation

Amplifier

Advertising