Table 12. ni 9203 calibration coefficients, Table 13. ni 9203 calibration equation information – National Instruments Deterministic Ethernet Expansion Chassis NI 9144 User Manual

Page 27

Advertising
background image

© National Instruments Corporation

27

NI 9144 User Guide and Specifications

The calibration data is stored in a U32 array, though each offset field should
be interpreted as a signed value.

Use the calibration coefficients with the following equation to generate
corrected data:

Table 12. NI 9203 Calibration Coefficients

Coefficient

Representation

Units

LSB Weight

Unsigned

pA/LSB

Offset

Signed

pA

Table 13. NI 9203 Calibration Equation Information

Term

Units

Definition

I

corrected

pA

Calibrated current

I

raw

bits

The raw code from the
NI 9203

I

expected0mA

bits

Expected code at 0 mA.
0 bits for 0–20 mA range.
32768 bits for ±20 mA
range

LSB

weight

pA/bit

Number of pA in one bit

I

offset

pA

Offset at 0 mA

I

corrected

I

raw

(

)

I

raw

I

expected 0mA

LSB

weight

I

offset

=

pA bits pA/bit pA

⎧ ⎪ ⎪ ⎨ ⎪ ⎪ ⎩

⎧ ⎪ ⎪ ⎪ ⎨ ⎪ ⎪ ⎪ ⎩

⎧ ⎪ ⎨ ⎪ ⎩

⎧ ⎨ ⎩