Table 29. ni 9217 conversion code, Table 30. ni 9217 scan list format – National Instruments Deterministic Ethernet Expansion Chassis NI 9144 User Manual

Page 37

Advertising
background image

© National Instruments Corporation

37

NI 9144 User Guide and Specifications

one conversion in the future, at time t+1. The conversion code is listed in
Table 29.

Note

The conversion rate for every channel must match the value of the conversion speed

control in 0x2002.

For example, the scan list entry 0x00000001FC indicates this scan stores at
address 1, and the next conversion is channel 2 at high-accuracy.

Table 30 contains the default scan list.

NI 9217 Calibration Data

The NI 9217 has four RTD channels that can measure 100

Ω RTD in 3-wire

and 4-wire mode. There is a 1 mA excitation current source per channel and
the module range is –500

Ω to 500 Ω. The resistance range specified in the

manual is 0 to 400

Ω. This range is tested and covers the temperature range

of –200 ºC to 850 ºC for the standard platinum RTD. The channel does not
read negative resistance.

Each channel has an associated LSB weight, which is the number of
Ω per bit, and an offset, which is the number of Ω per bit measured when
the inputs are grounded.

Table 29. NI 9217 Conversion Code

Bits

Field

7:3

Conversion rate: 0b11111 = 31,
High-Accuracy
0b00010 = 2, High-Speed

2:1

Channel number

0

Reserved

Table 30. NI 9217 Scan List Format

Index

Sub

Type

Value

0x2001

0

ARR:U32

5

1

4

2

0x0000 | 0xF8 | 0x02

3

0x0100 | 0xF8 | 0x04

4

0x0200 | 0xF8 | 0x06

5

0x0300 | 0xF8 | 0x00

Advertising