Table 21. ni 9213 conversion speed control, Table 22. ni 9213 error status field – National Instruments Deterministic Ethernet Expansion Chassis NI 9144 User Manual

Page 33

Advertising
background image

© National Instruments Corporation

33

NI 9144 User Guide and Specifications

NI 9213 Conversion Speed Control

The NI 9213 converts at two pre-defined rates, as controlled by the speed
control field.

Note

The conversion rate assumes that 18 channels are in the scan list.

NI 9213 Common Mode Error/Status

The error/status field is shown in Table 22.

NI 9213 Scan List

The scan list contains channels to convert, in order. The NI 9213 has
eighteen measurable channels:

0..15: Sixteen thermocouple channels (always measured in a ±78 mV
range)

16: One cold junction channel (always measured in a ±2.5 V range)

17: One auto zero channel (always measured in a ±78 mV range)

Table 21. NI 9213 Conversion Speed Control

Speed Control

Meaning

Conversion

2 (0x02)

High-Accuracy

55 ms/channel (.99 s total)

15 (0x0F)

High-Speed

740

μs/channel (13.32 ms total)

Table 22. NI 9213 Error Status Field

Bits

Field

31:8

Reserved

7

OT Error:
1= Open thermocouple was detected on the
last channel that was acquired

6

CMV Error:
1= Common mode voltage error was
detected on the last channel that was
acquired

5

GO Status: status of the gain override enable
bit

4

Reserved

3:0

Currently configured ADC data rate

Advertising