The calibration data is stored in a U32 array, though each Offset field
should be interpreted as a signed value.

The NI 9219 returns calibrated 24-bit (padded to 32-bits) AI data for all
modes and ranges.

To convert raw data into engineering units use the following formula:

y = m*x + b

b = offset based on range of the device (such as, –60 for ± 60 Volts
Voltage Measurement Range)

m = Gain (Full-Range/ (2

24

)): (such as, 120/(2

24

) for ± 60 Volts Voltage

Measurement Range)

14

3-Wire RTD

Pt1000

15

Pt100

16

Quarter-Bridge

350

Ω

17

120

Ω

18

Half-Bridge

±500 mV/V

19

Reserved

20

Full-Bridge

±62.5 mV/V

21

±7.8 mV/V

Table 35. NI 9219 Calibration Data

Coefficient

Representation

LSB Weight

Unsigned

Offset

Signed