NI 9264

NI 9264 Calibration Data

The NI 9263 has four channels with a nominal range of ±10.5 V. Each
channel has an associated LSB weight, which is the number of volts per bit,
and an offset, which is the number of volts per bit measured when the inputs
are grounded.

Note

LSB weight is referred to as Gain in the object dictionary.

The calibration data is stored in a U32 array, though each Offset field
(subindex 1, 3, 5, and so on) should be interpreted as a signed value.

Use the calibration coefficients with the following equation to generate
corrected data:

Table 58. NI 9264 Vendor Configuration Extensions

Index

Sub

Type

R/W

Description

0x2100

1

ARR:U32

Calibration = 16

2

R

Ch0 Gain

3

R

Ch1 Offset

...

...

...

8

R

Ch3 Gain

9

R

External Ch0
Offset

...

...

...

Table 59. NI 9264 Scan List Format

Coefficient

Representation

Units

LSB Weight

Unsigned

pV/LSB

Offset

Signed

nV

V

desired

Code

(

)

Code LSB

weight

nV

bits

---------

⎞ 10

9

V

nV

-------

Offset nV

(

) 10

9

V

nV

-------

+

=