Ni 9233, Table 42. ni 9233 vendor configuration extensions, Table 43. ni 9233 scan list format – National Instruments Deterministic Ethernet Expansion Chassis NI 9144 User Manual

Page 49

Advertising
background image

© National Instruments Corporation

49

NI 9144 User Guide and Specifications

NI 9233

As a DSA module, the NI 9233 does not synchronize to other modules and
free-runs at its own fixed rate.

NI 9233 Configure ADC

The NI 9233 (and NI 9229/9239) converts at various rates, controlled by
the field in the ADC conversion command.

0x65

Data Byte

0xA0

CRC value

Table 42. NI 9233 Vendor Configuration Extensions

Index

Sub

Type

R/W

Description

0x2002

0

U32

R/W

Configure ADC,
default = 0x0A

0x2100

0

ARR:U32

Calibration = 16

1

R

Ch0 Offset

2

R

Ch0 Gain

3

R

Ch1 Offset

...

...

...

8

R

Ch3 Gain

9

R

External Ch0
Offset

...

...

...

Table 43. NI 9233 Scan List Format

Bits

Field

7

Turbo Disable (NI 9233 only)

6:2

Clock Divisor

1:0

Clock Source = 2

Table 41. NI 9219 Configuration 1: Command Bytes (Continued)

Command Byte Value

Description

Advertising