Installing the Hardware

The DAQ Getting Started Guide contains non-software-specific
information about how to install PCI and PXI Express devices, as well as
accessories and cables.

Device Self-Calibration

NI recommends that you self-calibrate your S Series device after
installation and whenever the ambient temperature changes.
Self-calibration should be performed after the device has warmed up for the
recommended time period. Refer to the device specifications to find your
device warm-up time. This function measures the onboard reference
voltage of the device and adjusts the self-calibration constants to account
for any errors caused by short-term fluctuations in the environment.
Disconnect all external signals when you self-calibrate a device.

You can initiate self-calibration using Measurement & Automation
Explorer (MAX), by completing the following steps.

1.

Launch MAX.

2.

Select My System»Devices and Interfaces»NI-DAQmx
Devices»your device
.

3.

Initiate self-calibration using one of the following methods:

Click Self-Calibrate in the upper right corner of MAX.

Right-click the name of the device in the MAX configuration tree
and select Self-Calibrate from the drop-down menu.

Note

You can also programmatically self-calibrate your device with NI-DAQmx,

as described in Device Calibration in the NI-DAQmx Help or the LabVIEW Help in
version 8.0 or later.