Chapter 7

Counters

© National Instruments Corporation

7-27

NI 6124/6154 User Manual

Routing Counter n Source to an Output Terminal

You can route Counter n Source out to any PFI <0..15> or RTSI <0..7>
terminal. All PFIs are set to high-impedance at startup.

Counter n Gate Signal

The Counter n Gate signal can perform many different operations
depending on the application including starting and stopping the counter,
and saving the counter contents.

Routing a Signal to Counter n Gate

Each counter has independent input selectors for the Counter n Gate signal.
Any of the following signals can be routed to the Counter n Gate input:

RTSI <0..7>

PFI <0..15>

AI Reference Trigger (ai/ReferenceTrigger)

AI Start Trigger (ai/StartTrigger)

AI Sample Clock (ai/SampleClock)

AI Convert Clock (ai/ConvertClock)

AO Sample Clock (ao/SampleClock)

DI Sample Clock (di/SampleClock)

DO Sample Clock (do/SampleClock)

PXI_STAR

Change Detection Event

Analog Comparison Event

In addition, Counter 1 Internal Output or Counter 1 Source can be routed to
Counter 0 Gate. Counter 0 Internal Output or Counter 0 Source can be
routed to Counter 1 Gate.

Some of these options may not be available in some driver software.

Routing Counter n Gate to an Output Terminal

You can route Counter n Gate out to any PFI <0..15> or RTSI <0..7>
terminal. All PFIs are set to high-impedance at startup.