Chapter 4

Analog Input

NI 6124/6154 User Manual

4-8

ni.com

Figure 4-3. Differential Connection for Ground-Referenced Signals

on Non-Isolated Devices

Figure 4-4 shows how to connect a ground-referenced signal source to a
channel on an isolated S Series device.

Figure 4-4. Differential Connection for Ground-Referenced Signals

on Isolated Devices

Ground-

Referenced

Signal

Source

Common-

Mode Noise

and Ground

Potential

V

s

V

cm

+

+

I/O Connector

AI 0 Connections Shown

AI 0 GND

AI 0–

AI 0+

+

+

V

m

Measured

Voltage

Instrumentation

Amplifier

Non-Isolated S Series Device

Ground-

Referenced

Signal

Source

Common-

Mode Noise

and Ground

Potential

V

s

V

cm

+

+

I/O Connector

AI 0 Connections Shown

AI 0–

AI 0+

+

+

V

m

Measured

Voltage

Instrumentation

Amplifier

Isolated S Series Device

Digital

Isolators

Isolation

Barrier