Chapter 7

Counters

© National Instruments Corporation

7-11

NI 6124/6154 User Manual

Figure 7-12 illustrates this method. Another option would be to
measure the width of a known period instead of a known pulse.

Figure 7-12. Method 2

Method 3: Measure Large Range of Frequencies Using
Two Counters
—By using two counters, you can accurately measure
a signal that might be high or low frequency. This technique is called
reciprocal frequency measurement. In this method, you generate a long
pulse using the signal to measure. You then measure the long pulse
with a known timebase. The S Series device can measure this long
pulse more accurately than the faster input signal.

Pulse

F1

Pulse

F1

Gate

Source

1 2 …

N

Pulse-Width

Measurement

T =

N

F1

Frequency of F1 =

T

Width of

Pulse

N

Width of Pulse (T)