Figure 7-12. method 2 – National Instruments Network Device DAQ S User Manual
Page 83

Chapter 7
Counters
© National Instruments Corporation
7-11
Figure 7-12 illustrates this method. Another option would be to
measure the width of a known period instead of a known pulse.
Figure 7-12. Method 2
•
Method 3: Measure Large Range of Frequencies Using
Two Counters—By using two counters, you can accurately measure
a signal that might be high or low frequency. This technique is called
reciprocal frequency measurement. In this method, you generate a long
pulse using the signal to measure. You then measure the long pulse
with a known timebase. The S Series device can measure this long
pulse more accurately than the faster input signal.
Pulse
F1
Pulse
F1
Gate
Source
1 2 …
N
Pulse-Width
Measurement
T =
N
F1
Frequency of F1 =
T
Width of
Pulse
N
Width of Pulse (T)