Example (user defined)—connecting an instrument, 2 example (user defined)—connecting an instrument – YSI 5200 User Manual

Page 82

Advertising
background image

YSI Incorporated

80

5200 Recirculating System Monitor

2. Connect the switch to Auxiliary Input #1 or Auxiliary Input #2 using the 3-pole YSI 065942UL

Pluggable Socket Connector. See Figure 8-1.

Connect one wire to the ground (GND) terminal and the other to the In 1 or In 2 terminal.

NOTE: When connecting a switch to Auxiliary Input #2, set the input to Digital mode. See
Section 9.4.2 Input #2.

3. Set the 5200 to either Alarm on Open or to Alarm on Closed, depending on the type of switch

used. See Section 9.4.1 Input #1, Control.

IN 2

--

IN 1

--

GND--

Figure 8-1 Float Switch Wiring

8.2.2

Example (User Defined)—Connecting an Instrument

1. Connect the instrument to Auxiliary Input #2 using the 3-pole YSI 065942UL Pluggable Socket

Connector. See Figure 8-2.

2. Connect the negative wire to the ground (GND) terminal and the positive wire to the In 2

terminal.

3. For a 4–20mA input, connect a 250

Ω resistor between the ground (GND) and positive In 2

terminals. See Figure 8-2.

4. Set Auxiliary Input #2 to User Defined mode and select the correct scale (0–1 VDC, 0–5VDC, or

4–20mA). See Section,9.4.2 Input #2.