Troubleshooting

Section

6

6.3 SENSOR PERFORMANCE PROBLEMS


SYMPTOMS POSSIBLE

CAUSE

ACTION

REF

Probe not properly calibrated

Follow chlorine cal procedures

2.6.1

Membrane not properly installed or
may be punctured

Follow 6572 setup procedure

2.3.1

DO probe electrodes require
cleaning

Follow cleaning procedure. Use
6035 maint. kit

2.10.2

Water in probe connector

Dry connector; reinstall probe

2.3.2

Chlorine Charge too high >30
1. Anode tarnished

Follow cleaning procedure. Use
6035 maintenance kit

2.7
2.10.2

DO probe has been damaged

Replace 6572 probe

8/C

Free Chlorine reading
unstable or inaccurate

Internal failure

Return sonde for service

8

Probe requires cleaning

Follow probe cleaning procedure

2.10.2

Probe requires calibration

Follow cal procedures

2.6.1
3.7.4

pH probe reference junction has
dried out from improper storage.

Soak probe in tap water or buffer
until readings become stable

2.10.2

Water in probe connector

Dry connector; reinstall probe

2.3.2

Probe has been damaged

Replace probe

8/C

Calibration solutions out of spec or
contaminated with other solution

Use new calibration solutions

C

ORP fails Zobell check

Take into account temperature
dependence of Zobell solution
readings

5.3

pH, ORP, chloride,
ammonium, or nitrate
readings are unstable or
inaccurate. Error
messages appear during
calibration.

Internal failure

Return sonde for service

8

YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

6-

3