DISCRETE SAMPLING


NOTE THAT THIS FUNCTION WILL NORMALLY BE USED IN DRINKING
WATER STUDIES ONLY DURING THE BREAK-IN OF THE SENSOR AFTER
RECONDITIONING.

Select

1-Discrete sample from the Run menu. The Discrete sample menu will be displayed.

------------Discrete sample------------
1-Start sampling
2-Sample interval=4
3-File=
4-Site=
5-Open file

Select option (0 for previous menu):

Select

2-Sample Interval to type a number that represents the number of seconds between samples. The

maximum sample interval is 32767 seconds (9+ hours). The factory default sample interval is 4 seconds
and works best for most discrete sampling applications. See section 2.9.1 for more details.

Select

3-File to enter a filename with a maximum of 8 characters. This is the file to which you will log

readings.

If you started sampling without entering a filename, the default name NONAME1 will be assigned to your
file. Whenever you press 1-LOG last sample or 2-LOG ON/OFF from the menu, NONAME1 will be
opened during sampling. If this happens, and you want to restart the file with a different name, press
5-Close file and rename the file.

Select

4-Site to assign a site name with a maximum of 31 characters. This allows you to enter the name of

the site where you are sampling.

When you select

5-Open File, a file is opened and the number 5 changes to Close File. When you are

finished logging data to the file, press

5-Close File and number 5 changes back to Open File.


Now select

1-Start sampling to start discrete sampling.


After the initial sampling time interval has passed, (e.g. 4 seconds in the example above), sequential lines
of data (similar to that shown below) will appear on the screen.