2 logging gps readings – YSI 600DW-B Sonde User Manual

Page 161

Advertising
background image

650 MDS

Section 3

accomplished via the 6115 GPS cable that can be purchased as an optional accessory. Using the 6115 cable,
GPS readings can also be logged to the 650 memory along with sonde sensor readings taken at a particular
geographical location, which can then be relocated for subsequent studies.

The use of the GPS interface is easy, but the user must configure the hardware and software of the GPS unit as
follows before proceeding:

• Consult the user manual of the GPS unit and configure the instrument so that readings in the NMEA 0183

protocol are generated.

• Purchase from your GPS manufacturer a cable that connects at one end to the GPS unit and has at its

other end a male DE-9 connector for interface with the YSI 6115 GPS cable.


Once these requirements are met, proceed according to the following instructions, consulting Figure 2 in
Section 3.2.3 above for assistance:

• Connect the 6115 GPS cable to the 650 via the MS-8 connector.

• Connect the other MS-8 connector of the 6115 GPS cable to the MS-8 connector or your sonde cable.

• Attach the male DE-9 connector of the 6115 GPS cable to the mating female connector of your GPS unit.

• Turn on the 650 and make certain that appropriate GPS readings are displayed in the upper left corner of

the Status Bar as shown in the following display.
















3.8.2 LOGGING GPS READINGS

To log latitude and longitude readings from a user-supplied GPS unit, the following conditions must be
met:

• GPS readings can ONLY be logged to 650 memory – the system does NOT support logging to sonde

memory.

• The GPS readings must be logged to 650 memory from the 650 Run display along with sonde sensor

readings. This means that you must have a sonde attached to your 650 via the 6115 cable in order to
log GPS readings. Logging of GPS readings is not supported without sonde attachment even though
these readings can be viewed in the Status Bar with no sonde attached.

• If your want to log the GPS readings to 650 memory along with sonde sensor readings at a particular

location, you must make certain that the Store Lat and Long option is active in the Logging setup
menu of the 650.