Specifications Appendix

E

Internal logging
memory size:

384 kilobytes (150,000 individual parameter readings)

Power:

8 AA-size Alkaline Batteries or External 12 VDC

Battery Life:

Approximately 30 days at 20

°C at 15 minute logging intervals, a 40 second DO

warm up time and turbidity, chlorophyll, or rhodamine-WT active


SENSOR SPECIFICATIONS

The following are typical performance specifications for each sensor. Specifications do not include drift.

Temperature

Sensor Type...........Thermistor
Range.....................-5 to 45

°C

Accuracy................+/- 0.15

°C

Resolution..............0.01

°C


Conductivity

*

Sensor Type...........4 electrode cell with autoranging
Range.....................0 to 100 mS/cm
Accuracy................+/- 0.5% of reading + 0.001 mS/cm
Resolution..............0.001 mS/cm to 0.1 mS/cm (range dependent)

Free Chlorine, mg/L

Sensor Type.......... Rapid Pulse - Clark type, polarographic
Range.....................0 to 3 mg/L
Accuracy............... +/- 15 % of the reading or 0.05 mg/L, whichever is greater; to maintain accuracy
specification, flow must be at least 500 mL/min and pH should not change by more than
+/- 0.3 units if the mean pH is between 8.5 and 9.3. The sensor will not function properly

at pH values above 9.3

Resolution..............0.01 mg/L

Salinity
Sensor Type............Calculated from conductivity and temperature
Range......................0 to 70 ppt
Accuracy.................+/- 1.0% of reading or 0.1 ppt, whichever is greater
Resolution...............0.01 ppt

pH
Sensor Type...........Glass combination electrode
Range.....................0 to 14 units
Accuracy................+/- 0.2 units
Resolution..............0.01 units
Depth……………200 meters

ORP

Sensor type............ Platinum button

Range.....................-999 to +999 mV

* Report outputs of specific conductance (conductivity corrected to 25 C), resistivity, and total dissolved solids are also
provided. These values are automatically calculated from conductivity according to algorithms found in Standard
Methods for the Examination of Water and Wastewater
(Ed 1989).

YSI Incorporated Drinking Water Monitoring Systems Operations Manual

E-2