YSI 600DW-B Sonde User Manual

Page 95

Advertising
background image

Sondes

Section

2

The following table provides the default value, operating range, and comments relative to the calibration
constants. Error messages will appear during calibration if values are outside the indicated operating range
unless the designation is “not checked”.

Parameter

Default

Operating range

Comments

Cond:

5

4

to

6

Traditional

cell

constant

Cl2 gain:

1

0.5 to 2.0

mV

offset:

0.0

-100

to

100

pH offset:

0.0

-400 to 400

pH gain:

-5.0583

-6.07 to -4.22

NH4

J

51.2

Not

checked

NH4

S

0.195

0.15

to

0.217

NH4

A

1.092

Not

checked

NO3

J

99.5

Not

checked

NO3 S

-0.195

-0.217 to -0.15

NO3

A

2.543

Not

checked

Cl J

99.5

Not checked

Cl S

-0.195

-0.217 to -0.15

Cl A

2.543

Not checked

Turb Offset

0

-10 to 10

Turb A1

500

0.6 to 1.5

Range is ratio of M1 to A1

Turb M1

500

Turb A2

1000

0.6 to 1.5

Range is ratio of (M2-M1) to (A2-A1)

Turb M2

1000

Chl Offset

0

-30 to 20

Chl A1

500

0.6 to 1.5

Range is ratio of M1 to A1

Chl M1

500

Chl A2

1000

0.6 to 1.5

Range is ratio of (M2-M1) to (A2-A1)

Chl M2

1000

Rhod Offset

0

-10 to 10

Rhod A1

500

0.6 to 1.5

Range ratio of M1 to A1

Rhod M1

500

Rhod A2

1000

0.6 to 1.5

Range is ratio of (M2-M1) to (A2-A1)

Rhod M2

1000

To reset a calibration cell constant, access the sonde Calibrate menu. Then select the sensor and type
“UNCAL” instead of the value. This action will change that sensors calibration constants back to the
factory default.

From the Advanced menu, select

2-Setup to display miscellaneous options. Type the appropriate number to

activate/deactivate any of the displayed features.

YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-88