YSI 600DW-B Sonde User Manual

Page 32

Advertising
background image

Sondes

Section

2

studies,

Auto sleep RS232 should usually be “on”. This is described in detail in Section 2.9, Sonde Menu.

When this setup is verified, press

Esc or 0 to return to the Advanced menu.


Select

3-Sensor from the Advanced menu and make certain that the entries are identical to those shown

below.

------------Advanced sensor-----------
1-TDS constant=0.65
2-Cl2 temp co %/C=2.5
3-Cl2 warm up sec=60
4-( )Wait for Cl2
5-Wipes=1
6-Wipe int=1
7-SDI12-M/wipe=1
8-Turb temp co %/C=0.3
9-(*)Turb spike filter
A-Chl temp co %/C=0
B-( )Chl spike filter

Select option (0 for previous menu):


The display under

3-Sensor may be different from the one shown in the example above, depending on the

sensors that are installed in your unit. For example, if you do not have a chlorophyll probe, the last two
entries (which are relevant only to chlorophyll) will not appear.

When this setup is verified, press

Esc or 0 to return to the Advanced menu. For a detailed explanation of

the choices in the Advanced menu, see

Section 2.9.8, Advanced. Press Esc or 0 to back up to the Main

menu.

------------------Main-----------------
1-Run 5-System
2-Calibrate 6-Report
3-File 7-Sensor
4-Status 8-Advanced

Select option (0 for previous menu):


The sonde software is now set up and ready to calibrate and run.

YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-25