Optic chlorophyll – YSI 600DW-B Sonde User Manual

Page 80

Advertising
background image

Sondes

Section

2

Following the instructions on the screen, place the sonde in the second turbidity standard, input the correct
turbidity value in NTU, press

Enter, and view the stabilization of the values on the screen in real-time.

After the readings have stabilized, press

Enter to confirm the calibration. Then, as instructed, press any

key to return to the Calibrate

menu.


Select the

3-point option for maximum accuracy over the entire range of 0 to 1000 NTU. As for the 2-

point procedure, one of the standards must be 0 NTU. Because of the linearity characteristics of the
sensors, we recommend that the other two standards have turbidity values of 100 and 1000 NTU.
However, the user can select any values that are deemed appropriate. The procedure for this calibration is
the same as for a 2-point calibration, but the software will prompt you to place the sonde in the additional
solution to complete the 3-point procedure.

For all turbidity calibration procedures, be certain that the standard and sensor are thermally equilibrated
prior to proceeding with the calibration.

For further information related to setting up, calibrating and running turbidity measurements, refer to
Section 5.11.

OPTIC CHLOROPHYLL


When selecting

10-Optic Chlorophyll, there will be a choice of zeroing the full scale range of the

fluorescence sensor (

Fluor Zero), or calibrating with actual chlorophyll standards (µg/L 1-point, 2-point,

or 3-point).

If you have selected

Fluor %FS as a parameter in Report Setup, the sonde will report only relative values

of fluorescence in the sample being measured. These values could then be converted into actual chlorophyll
concentrations in µg/L by using a post-calibration procedure, after the chlorophyll content of grab-samples
taken during a sampling or monitoring study has been analyzed in a laboratory. This determination can
involve conducting the extractive analysis procedure described for chlorophyll in Methods for the
Examination of Water and Wastewater
or by carrying out an in situ measurement of chlorophyll using a
commercial benchtop fluorometer.

The

Fluor zero option is used to zero the fluorescence probe in a medium that is chlorophyll-free. Place

the sonde in clear water, and input 0 at the screen prompt. Press

Enter and the screen will display real-

time readings that will allow you to determine when the fluorescence readings have stabilized. Press

Enter

after the readings have stabilized to confirm the calibration and zero the sensor. Then, press any key to
return to the Calibrate menu.

If you select

Chl µg/L in the initial calibration routine, there will be a choice of 1-point, 2-point, or 3-point

options. The 1-point selection is normally used to zero the fluorescence probe in a medium that is
chlorophyll-free. If you use this method, you will either choose to utilize the default sensitivity for
chlorophyll in the sonde software or to update a previous multipoint calibration. Usually you will place the
sonde in clear water, and input 0 µg/L at the screen prompt. After pressing

Enter the screen will display

real-time readings allowing you to determine when the chlorophyll readings have stabilized. Press

Enter

after the readings have stabilized to confirm the calibration and zero the sensor. Then, as instructed, press
any key to return to the Calibrate menu.

Note: For the 2-point and 3-point calibrations described below, standards of known fluorescence are
required. Two general types of standards can be used: (a) phytoplankton suspensions of known chlorophyll
content, and (b) dye solutions whose fluorescence can be correlated to that of chlorophyll. The user is
responsible for determining the chlorophyll content of phytoplankton suspensions, either by employing the
extractive analysis procedure described in Standard Methods for the Examination of Water and
Wastewater,
or by analyzing the suspension in situ using a laboratory fluorometer. See Section 5,
Principles of Operation of this manual for more information about chlorophyll standards.

YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-73