YSI 600DW-B Sonde User Manual

Page 151

Advertising
background image

650 MDS

Section 3




NOTE: If you want to return to the Edit site list display for changes before logging, use the RIGHT
ARROW KEY to highlight the Configure selection and press Enter. The up/down arrows will only scroll
within the Site list.

The data for two different sites has now been logged to the single file (BLUELAKE) that can be either
viewed manually using the 650 View file menu or uploaded to EcoWatch for Windows from the 650
File|Upload to PC
menu. The Site Number (“1” for the WEST site and “2” for the EAST site) will be
stored with each data point and will be thus be available for identifying where the data was taken. The file
BLUELAKE can be examined by use of the 650 File selection. In the 650 File menu, highlight Directory
and press Enter as shown below. Then press Enter with the cursor on the BLUELAKE file to show the
file details.














Note that there is NO Site name listed for the file BLUELAKE since data from more than on site was
logged to it. Now highlight the View file selection and press Enter to display the data in BLUELAKE as
shown below.











Note that the data from two sites is identified by the Site Number attached to each data record. To index
these numbers to the appropriate locations, the user must consult the Site list by activating Edit site list
from the 650 Logging setup menu. The correlation between site names and numbers is found here,
indicating that Site 1 is WEST and Site 2 is EAST.








YSI, Incorporated Drinking Water Monitoring Systems Operations Manual

3-42

Advertising
This manual is related to the following products: