YSI 600DW-B Sonde User Manual

Page 99

Advertising
background image

Sondes

Section

2

CAUTION: If Wipe Int is set to zero, then no wiping will occur either in Discrete or
Unattended Sampling. Make certain that Wipe Int is set to some finite value prior to
setting up an Unattended study or no automatic cleaning will occur.


SDI12-M/Wipe=1 This is the number of wiping cycles when the sonde is in SDI12 mode. The wiper for

the 6026 and 6136 turbidity, 6025 chlorophyll, and 6130 rhodamine WT sensors will
automatically wipe each time this many SDI12 “M” commands have been issued. If this
value is set to zero, then no automatic wiping will occur. This item will only appear if a
turbidity, chlorophyll, or rhodamine WT sensor is enabled in the Sensors menu.


Turb temp co %/C=0.3 This entry sets the coefficient for the temperature compensation of turbidity

readings from the 6026 and 6136 sensors. The default values of 0.3 (6026) and 0.6
(6136) should not be changed by the user without consulting YSI Customer Service. This
item will only appear if a turbidity sensor is enabled in the Sensor menu.


(* ) Turb Spi
ke Filter When this item is activated, the output of the turbidity sensor is mathematically

processed to minimize the effect of unusual (or “bad”) readings on the overall data
presentation. In most cases, these “spike” events are the result of the chance passage of
large suspended particles across the probe optics just at the time a reading is taken.
Activation of this option generally results in a better display of the “average” turbidity of
the water under examination and its use is recommended for most sampling and
unattended applications. This item will only appear if a turbidity sensor is enabled in the
Sensors menu.


Chl temp co %/C=0.0 This entry sets the coefficient for the temperature compensation of chlorophyll

readings from the 6025 sensor. The default value of zero should only be changed by the
user after establishing the temperature compensation factor for the phytoplankton sample
in question. See Section 5.12 of this manual for more information. This item will only
appear if a chlorophyll sensor is enabled in the Sensors menu.



From the Advanced menu, select

4-Data filter to display filtering options. Type the appropriate number to

activate/deactivate any of the displayed features.

Sondes with

no optical (turbidity, chlorophyll, or rhodamine WT) probes enabled will display the

following menu.


------------Data filter setup----------
1-(*)Enabled
2-( )Wait for filter
3-Time constant=4
4-Threshold=0.001

Select option (0 for previous menu):

If any optical probe (turbidity, chlorophyll, or rhodamine WT)

is enabled, then the menu will appear as

follows.

YSI Incorporated

Drinking Water Monitoring Systems Operations Manual

2-92

Advertising
This manual is related to the following products: